Kipp & Zonen: SMP12, Class A pyranometer
Kipp & Zonen SMP12 Class A pyranometer
The new SMP12 is pioneering fast response spectrally flat Class A pyranometer that combines solid-state dome heating preventing dew and frost, with no moving parts, and best-in-class surge protection to maximize accuracy and minimize maintenance.
The SMP12 combines proven sensor and diffuser technology from the RaZON+ PH1 and PR1 radiometers with an enhanced version of the Smart interface and the low maintenance of the SMP10.
The SMP12 has an internal drying cartridge that will last for at least 10 years if the housing is not opened. This further reduces maintenance significantly.
The SMP12 feature set includes:
- Solid-state dome heating for dew and frost prevention.
- Multi-axis tilt sensor to enable remote monitoring of the pyranometer mounting and orientation.
- Internal humidity and temperature monitoring to provide confidence in the integrity of the instruments internal functionality.
Benefits
- ISO / IEC classification
ISO 9060 spectrally flat fast response Class A and IEC 61724-1 Class A monitoring, with ISO / IEC 17025 calibration. - Remote tilt angle and internal humidity monitoring
Tilt sensor to enable remote checks of alignment and humidity monitor provides confidence that the sealing has not been compromised - Minimized maintenance
No desiccant change for 10 years, no moving parts, best MTBF with 5 year warranty, RS-485 connectivity and Modbus® RTU protocol for easy integration with data loggers and SCADA systems - Pioneering Class A pyranometer with integrated solid state heating and no moving parts
Heated dome to prevent dew and frost to achieve all-weather performance - Enhanced surge protection
Best-in class surge protection conform to EN 61000-6-2 to minimize failure risks, and maintain performance - Smart Interface
2-wire RS-485 Modbus® RTU interface
Tilleggsinformasjon
Grensesnitt | |
---|---|
Måleparameter | |
Miljø |
Dokumentasjon
Spesifikasjoner
Cable Length | 33, 82, 164 ft (10, 25, 50 m) |
Classification | Fast response Spectrally Flat Class A (ISO 9060:2018) |
Digital Outputs | Modbus® RTU over 2-wire RS-485 |
Directional Response | ˂ ±10 W/m² (up to 80° with 1000 W/m² beam) |
Drying Cartridge and Maintenance Interval | Internal, lasts for 10 years, replaced with every recalibration |
IP Rating | IP67 |
Material Enclosures | Aluminum, anodized |
Maximum Irradiance | 2000 W/m² (Max.) |
Non-linearity | ˂ ±0.2% (100 to 1000 W/m²) |
Non-stability | ˂ ±0.5 % (change/year) |
Operating Humidity | 0 to 100 % |
Operating Temperature Range | -40 to +70 °C (Storage: -40 to +80 °C) |
Response Time | ˂ 0.15 s (63%), 0.5 s (95%) |
Spectral Accuracy | 280 to 3000 nm |
Spectral Range (50% points) | 285 to 2750 nm |
Temperature Dependence of Sensitivity | +14 to +104 °F (-10 to +40 °C): ˂ ±1%, -40 to +158 °F (-40 to +70 °C): ˂ ±2% |
Total Zero Offset | ˂ ±3 W/m² (including A & B) |
Weight | 1.1 lb (500 g) |
Zero offset A | ˂ ±1 W/m² |
Zero offset B | ˂ ±1.5 W/m² |
Technical Description
The SMP12 has a 2-wire RS-485 Modbus® RTU interface but, unlike the other SMP models, it does not have an analog output. The characteristics of the thermopile detector provide a fast response time of less than half a second. The wide power supply range from 10 to 30 VDC makes integration into solar energy plant performance monitoring stations easy. SMP12 has class-leading electrical protection in the form of 2-stage surge and reverse polarity protection on the power input and 3-stage surge protection on the RS-485 data lines.
Like all SMP series pyranometers, the calibrated sensitivity is stored in memory and can be accessed via the Modbus® interface and the measured irradiance value is output via Modbus® in W/m². This makes it easy and simple to interchange different instruments when recalibration or repair is required.
SmartExplorer Windows™ software provides: data logging on a PC, displaying of data, changing Modbus® communication settings, configuring of the Modbus® addresses – and more. SmartExplorer is free and can be found under the ‘Downloads’ tab of this webpage.