HOBO Pro v2 Temp/RH

Legg til i ønskeliste
Legg til i ønskeliste
Produktnummer: U23-001A Kategori:

The HOBO Pro v2 Temperature/Relative Humidity data logger is a weatherproof data logger with built-in temperature and relative humidity sensors. The Relative Humidity sensor provides superior durability in humid environments and is user-replaceable.

Diagram of data retrieval from U23 data logger using optic base station.

Highlighted Features

  • Weatherproof housing for use in outdoor or condensing environments
  • High accuracy
  • Replaceable Relative Humidity sensor provides fast response and superior recovery from condensing conditions
  • Small-diameter external sensor versions for measurements in tight spaces
  • Optic USB interface for fast and reliable download
  • Compatible with HOBOware and HOBOware Pro software for logger setup, graphing and analysis (HOBOware Pro is required when using the U-DTW-1 Waterproof shuttle)

HOBO data loggers require HOBOware software and a USB cable. (Note: A USB cable is included w/HOBOware Pro CD or USB software purchase). HOBOware Pro is required when using the U-DTW-1 Waterproof shuttle. See compatible items below or build your kit above.

Tilleggsinformasjon

Måleparameter

, ,

Miljø

Spesifikasjoner

Temperature Sensor
Operation Range U23-001A internal sensor: -40 to 70°C (-40 to 158°F)
U23-002A external temperature sensor: -40 to 70°C (-40 to 158°F)
U23-003 and U23-004 external sensors: -40 to 100°C (-40 to 212°F), with tip and cable immersion in fresh water up to 50°C (122°F) for one year
Accuracy U23-001A and U23-002A: ±0.25°C from -40 to 0°C (±0.45 from -40 to 32°F), ±0.2°C from 0 to 70°C (±0.36 from 32 to 158°F)
U23-003 and U23-004: ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F); see Plot A
Resolution U23-001A and U23-002A: 0.04°C (0.072°F)
U23-003 and U23-004: 0.02°C at 25°C (0.04°F at 77°F); see Plot A
Response Time
(Typical to 90%)
U23-001A internal sensor: 10 minutes in air moving 1 m/sec
U23-002A external temperature sensor: 3 minutes, 45 seconds in air moving 1 m/sec
U23-003 and U23-004 external sensors: 3 minutes in air moving
1 m/sec; 30 seconds in stirred water
Drift U23-001A and U23-002A: <0.01°C (0.018°F) per year
U23-003 and U23-004: <0.1°C (0.18°F) per year
Relative Humidity Sensor (U23-001A, U23-002A only)
Operation Range 0 to 100% RH, -40° to 70°C (-40° to 158°F)
Exposure to conditions below -20°C (-4°F) or above 95% RH may temporarily increase the maximum RH sensor error by an additional 1%
Accuracy ±2.5% from 10% to 90% RH typical to a maximum of ±3.5% including hysteresis at 25°C (77°F); below 10% and above 90% ±5% typical
Resolution 0.05%
Response Time
(Typical to 90%)
U23-001A: 40 minutes in air moving 1 m/sec with protective cap
U23-002A: 15 seconds in air moving 1 m/sec
Drift <1% per year typical
Logger
Operation Range -40° to 70°C (-40° to 158°F)
Real-time Clock ±1 minute per month 0° to 50°C (32° to 122°F)
Battery 1/2 AA, 3.6 Volt lithium, user-replaceable (HP-B)
Battery Life (Typical Use) 3 years with 1 minute or greater logging interval
Memory (Non-Volatile) 64K bytes memory (approx. 21,000 temperature and RH measurements)
Materials All models: ASA styrene polymer housing and mounting clamp; polypropylene protective cap; Buna-N o-ring(s); U23-001A,
U23-002A only: ASA styrene polymer RH sensor cap; modified hydrophobic polyethersulfone membrane
Cables U23-001A: No cables
U23-002A: One 184 cm (6 ft) PVC cable; sensor diameter 1 cm (0.38 in.)
U23-003: Two 184 cm (6 ft) PVC cables; sensor diameter 0.5 cm (0.20 in.)
U23-004: One 184 cm (6 ft) PVC cable; sensor diameter 0.5 cm (0.20 in.)
Environmental Rating Electronics housing is NEMA 6P equivalent (tolerant of brief submergence);
Units with RH sensors are NEMA 4 equivalent (splash-resistant)
Launch Modes Immediate start; delayed start
Logging Interval Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours
Offload Modes Offload while logging; stop and offload
Battery Indication Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
Weight U23-001A: 57 g (1.5 oz); U23-002A: 118 g (3.1 oz); U23-003: 138 g (3.7 oz); U23-004: 102 g (2.7 oz)
Dimensions Housing measures 10.2 × 3.8 cm (4.0 × 1.5 in.)
NIST Certificate Temperature certificate available for additional charge
 

Image
Description: ce

 

The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).